7-l-4.
If the reading exceeds tolerance,
adjust the
tilt
of condenser
C2-CH carrier by loosen-
ing 4 clamping
screws
(AB3
x
lOSA),
tilting
the entire assembly accordingly
and
tightening
the 4 clamping screws.
7-2. Adjustment
of Y (vertical)
direction
7-2-l. Screw jig (SKN0004)
into the nosepiece.
7-2-2. Rotate the jig (SKN0004)
to bring it paral-
lel to the Y-direction.
Bring stage upwards
until
bottom
surface of jig contacts
stage
surface,
-4o-