Olympus CK40M FRAME Manual de reparación - Página 9
Navegue en línea o descargue pdf Manual de reparación para Microscopio Olympus CK40M FRAME. Olympus CK40M FRAME 38 páginas. Compact inverted metallurgical microscope
CK40M
(1)
Reflection protector (AD378100) mounting position
Place the reflection protector on
the plate and stick it, aligning the
right and left widths of reflection
protector to the center hole.
(2)
Half mirror (LP148000) mounting position
Set the positioning jig (CK40MKC01) on the half
mirror mount. Determine the half mirror position by
pushing the half mirror to the CK40MKC01 and
apply silicone adhesive (OT1873) to the four corners
(bevelled parts) of the half mirror.
The range of adhesive coating should be at least
7 mm per location.
Half mirror
(LP148000)
Fix the half mirror using jigs
(CK40MKC01, K2347).
B. DISASSEMBLY AND ASSEMBLY PROCEDURES
Longer
Half mirror (LP148000)
Adhesive (4 locations)
B-2
Reflection protector (AD378100)
Double-coated adhesive tape
Plate (AD377800)
Half mirror mount
(AD377700)
Make sure that the coated
surface faces down.
Positioning jig
(CK40MKC01)
V-block (K2347)
Positioning jig
(CK40MKC01)