Olympus CX33 Manual de instrucciones - Página 29

Navegue en línea o descargue pdf Manual de instrucciones para Microscopio Olympus CX33. Olympus CX33 40 páginas. Biological microscope

Olympus CX33 Manual de instrucciones
7

Specifications

Item
Optical system
Infinity optical system
Microscope frame
CX33RTFS2 / CX33LTFS2
Illumination system
Built-in LED light source
Microscope frame (rated input power): 5 V 0.85 A
AC adapter (rated input power): 100-240 V
AC adapter (rated output power): 5 V
Focusing mechanism
Stage height adjustment mechanism
Movement distance per scale of fine focusing knob: 2.5µm
Movement distance per rotation of fine focusing knob: 0.3mm
Movable range: 15mm (Upward by 1.5 mm and downward by 13.5mm from focal position)
Equipped with pre-focusing lever; Tension adjustable coarse focusing knob
Revolving nosepiece
Fixed 4-hole revolving nosepiece
Tube
Type
Field number
Tube tilting angle 30°
Interpupillary
distance
adjustable range
Light path
selection
Stage
Movable range
Specimen holder
Condenser
Dimension / Weight
211 (W) x 390 (D) x 430 (H) mm / Approx. 7.0 kg (Not including AC adapter)
· Indoor use
Operating environment
· Altitude: Max. 2000 meters
· Ambient temperature: 5 to 40 °C (41 to 104 °F)
· Humidity: Max. 80% (31 °C or less) (without condensation)
· Supply voltage fluctuation: ±10 %
· Pollution degree: 2 (in accordance with IEC60664-1)
· Installation (overvoltage) category: II (in accordance with IEC60664-1)
· Ambient temperature: -25 to 65°C (41 to 104 °F)
Transportation / storage
· Humidity: 0% to 90% (without condensation)
environments
Trinocular tube (fixed type)
20 (when using standard 10X eyepieces)
48 to 75mm
2 level selection: Eyepiece 100% or Camera100%
X direction: 76mm
Y direction: 52mm
Possible to hold 2 slide glasses
Option
CX3-SHP (Specimen hold plate)
Model type
Abbe condenser
Numerical
1.25 (when immersed with oil)
aperture
Aperture
Illumination aperture diaphragm blade, manual open/close type
diaphragm
In case of over 31 °C (88 °F), the humidity in operating environment is decreased linearly
through 70% at 34 °C (93 °F), 60% at 37 °C (99 °F), and to 50% at 40 °C (104 °F).
Specification
50-60 Hz 0.4 A
2.5 A
CX33LTFS2
Possible to hold 1 slide glass
CX3-SHP (Specimen hold plate)
CX3-HLDT (Specimen holder)
CX33
CX33RTFS2
25