Olympus FV3000 Manual de instrucciones - Página 6
Navegue en línea o descargue pdf Manual de instrucciones para Microscopio Olympus FV3000. Olympus FV3000 38 páginas.
También para Olympus FV3000: Manual de seguridad (16 páginas), Manual de instrucciones (5 páginas)
XY Image Acquisition(1)
②
⑩
④
4
①
⑧
Reducing noises by
Averaging.
Adjusting Confocal
Aperture.
⑨
Laser Intensity
Sensitivity
⑤
③
⑦
① Select "VBF" in the [PMT Setting] Tool Window.
Assigning the detector to channel
② Press the
Tool Window to open the [Dye & Detector Select]
dialog box.
③ Press the
Dye.
④ Double-click the name of the fluorescence dye to
observe.
Virtual Channel Scan
This mode allows to acquire multiple CH image
whose channel number is larger than the
number of detector.
⑤ When registering a fluorescence dye, TD channel
is registered automatically. If you don't need,
double-click the TD channel to cancel it.
⑥ If you want to observe with multiple phases using
the virtual channel scan, press the
to add the phase. And drag&drop the selected
"Dye" to the observation channel list of the phase
you want to add.
⑦ After setting all channels, press the [OK] button.
Adjusting the live image
⑧ If you want to acquire multiple channel image,
select Sequential scan "Line" in the [PMT Setting]
Tool Window.
⑨ Press the
button on the [Live] window.
Adjust focus and set Laser Intensity(%),
Sensitivity(HV), Gain and Offset on [PMT setting]
Tool Window.
⑩ If you use virtual channel scan, press
to switch phases. On both phases, adjust focus
and set Laser Intensity(%), Sensitivity(HV), Gain
and Offset.
【Hi-Lo Mode】
pixels of intensity 4095 are shown in red.
Pixels of intensity 0 are shown in red.
button on [PMT Setting]
button to reset the Assigned
button
u5