Olympus SZ-ADD Manual de instrucciones - Página 12
Navegue en línea o descargue pdf Manual de instrucciones para Microscopio Olympus SZ-ADD. Olympus SZ-ADD 16 páginas. Sz2-ila transmitted illumination attachment
@
²
Fig. 2
Fig. 3
4
Fig. 4
9
1
}The applicable lamp bulb model is the 12V100WHAL-L (PHILIPS 7724)
1. Fully loosen the clamping screw @ at the top of the lamp housing using
2. Lift up the lamp housing ² to remove it.
3. Tilt the bulb socket ³ by 90° in the direction of the arrow.
4. While pushing down the bulb clamping lever | down, hold the halogen
³
5. Fit the lamp housing from up and tighten the clamping screw @ by
5
6
7
}Refer to the instruction manual for the LSGA.
Attaching the Halogen Bulb
With the U-LH100L-3
halogen bulb.
the Allen screwdriver provided with the microscope body.
bulb 5 with gloves or a piece of gauze, insert the bulb pins 6 straight
and fully into the pin holes 7 on the lamp socket.
Then return the lamp clamping lever gently back to the original position
to clamp the bulb.
To prevent reduced bulb life or cracking, do not touch the bulb with
bare hands. If fingerprints are accidentally left on the bulb, wipe the
bulb with a soft cloth.
applying downward pressure. (Fig. 2)
Caution for Bulb Replacement During or Right After Use
The bulb, lamp housing and areas near these will be extremely hot
during and right after use.
Set the main switch to "
the wall outlet, then allow the old bulb and lamp socket to cool
before replacing the bulb with a new of the designated type.
With the LSGA
(Figs. 2 to 4)
" (OFF), disconnect the power cord from