from your microscope
due to unfamiliarity
with its use, the table below may provide
some guidance
The
interpupillary
distance
is
not
adjusted
co.rrectly.
1. Incomplete
binocular
Vision
Correct the interpupillary
distance.
Diopter adjustment
is incomplete.
Complete
diopter
adjustment.
Replace
and
mount
the same
ey
pieces.
The
right
and
left
eyepieces
are
different.
Stains or dust have accumulated
on
the specimen.
Clean thoroughly.
2. Stains or dust are observed
on
the field of view.
Stains or dust have accumulated
on
the eyepieces.
Clean thoroughly
3. Unclear image
Clean thoroughly
Stains or dust have accumulated
on
the objective.
Diopter adjustment
of the eyepieces
is
incomplete.
Complete
diopter
adjustment.
4. Image blurs when zoomed.
Focus adjustment
is incomplete.
Complete
focus adjustment
The focusing
knob tension adjustment
is too tight.
Loosen appropriately.
5. The focusing
knob is stiff.
6. Poor focus during Observation
due to unintentional
lowering
of the zoom microscope
body.
The
focusing
knob
tension
is too
loose.
Tighten
appropriately
7. The right eyepiece
field or the
photomicrographic
equipment
viewer field is partially tut off.
The light path selector lever is between
settings.
Set to the correct
Position.
8. The
Video
monitor
image
goes excessively
out of focus
when
zoomed.
The
tube
length
adjustment
is in-
complete.
Complete
tube length adjustment.