Olympus SZX-2RE Manual de instrucciones - Página 23

Navegue en línea o descargue pdf Manual de instrucciones para Microscopio Olympus SZX-2RE. Olympus SZX-2RE 34 páginas. Research stereomicroscope system

Olympus SZX-2RE Manual de instrucciones

OPERATION OF OTHER ATTACHMENTS

9-1 Eyepoint Adjuster SZX-EPA

1
External View
2
Assembly
19
Observation tube clamping screw
@
²
Fig. 27
1. Using the Allen screwdriver provided with the SZX microscope body,
remove the observation tube @.
2. Mount the eyepoint adjuster ² to the place where the observation tube
has been mounted.
3. Mount the observation tube (removed in step 1 above) onto the eyepoint
adjuster.
This allows you to increase the eyepoint height by 40 mm. Up to two
eyepoint adjusters can be mounted provided that no other intermediate
attachment is used.
(Fig. 27)