Using the Joint Plates
NOTES
24
@
Fig. 6
Fig. 7
The joint plate @ can be attached and locked between the slider knob
and slider as shown in Fig. 6. The joint plates should be attached on
both ends of the filter slider.
By locking with the joint plates, you can switch the barrier and excitation
filters together as a set.
· When the 6-position filter slider near the rear panel is used, avoid using
the interference type or color glass type filters.
This is because the 6-position filter slider near the rear panel is one
of the positions where the energy from the light source is concen-
trated. When an interference type or color glass type filter is mounted
in it, the filter interference film may peel off or the color glass may be
damaged.
· Make sure that the 6-position filter slider is set to a click position.
· For safety, insert the provided light shield plates in the unused filter posi-
tions.