Olympus APU-300 Manual de consulta rápida - Página 4
Navegue en línea o descargue pdf Manual de consulta rápida para Equipamiento médico Olympus APU-300. Olympus APU-300 4 páginas.
Monopolar Argon Plasma Coagulation (MAPC)*
Procedure
Treatment of adenoma
remnants
Diffuse bleeding, vascular
malformations and
angiodysplasia in the
small bowel and colon
Barrett's esophagus
Gastric antral vascular
ectasia (GAVE)
Respiratory Procedures
Procedure
Less extended lesion,
bleeding
Polypoid lesion
(intraluminal mass)
Web-like stenosis
Tissue ablation
Please note
The operation settings are examples based on previous general clinical uses of the device. The surgeon is solely responsible for
selecting an appropriate mode and an appropriate output power level. The appropriate settings depend on the instrument to be
used and on the conditions of the tissue to be treated.
* Olympus recommends to always stick to the default argon flow rate of the respective argon probe.
Instrument or technique
During activation, keep a
distance of 2 - 5 mm between
argon probe's tip and tissue.
During activation, keep a
distance of 2 - 5 mm between
argon probe's tip and tissue.
During activation, keep a
distance of 2 - 5 mm between
argon probe's tip and tissue.
During activation, keep a
distance of 2 - 5 mm between
argon probe's tip and tissue.
Instrument or technique
CD-6C-1
SD-7C-1
KD-31C-1
During activation, keep a
distance of 2 - 5 mm between
argon probe's tip and tissue.
Manufacturer:
Olympus Winter & Ibe GmbH
Kuehnstr. 61
22045 Hamburg, Germany
Mode
Power
PulsedArgon Fast
30
SmartArgon
120
PulsedArgon Fast
30
PulsedArgon Fast
30
Mode
Power
SoftCoag
40
PulseCut Slow
50
PulseCut Slow
40
ForcedCoag
40
ForcedArgon
35
W9218035_00, en
Eff ect
n/a
2 - 4
n/a
n/a
Eff ect
3
2
2
3
n/a
2018-10-11