Olympus EMPOWER EMP-H35S Manual de consulta rápida - Página 5
Navegue en línea o descargue pdf Manual de consulta rápida para Equipamiento médico Olympus EMPOWER EMP-H35S. Olympus EMPOWER EMP-H35S 10 páginas. Yag lasers
Connect the laser fiber
The fiber is connected to the device
n
through the fiber port on the front panel.
The fiber connector has an additional
n
ring which facilitates the clamping of
the fiber to the connector on the device.
Furthermore, this ring enables the
automatic detection of fiber status
(present/absent), its diameter and its
type (single use/reusable). If the fiber
is not connected to the device, an error
message is displayed when the device
is switched on.
On the H65 and H100, a dedicated sensor
n
recognizes the presence of the user
hand, opening the external protection
shutter of the fiber connector.
Standby/Ready
When the user presses the Ready/Standby buttons, the system will be set in the Ready or Standby mode,
n
respectively. The button on the footswitch, or the Ready/Standby display area can be used interchangeably.
In Standby mode, the laser is not firing and the system cannot emit energy.
n
In the Ready mode, the laser is firing and ready to emit energy.
n
H35 Standby and Ready screens
L A S E R O P E R A T I O N
H65 & H100 Standby and Ready screens
E M P - H 3 5 S , E M P - H 65 S & EM P -H 100 S QUI C K REF ER ENC E G UID E
5