Output Check 2 (Electrosurgical system)
After proper operation of the PSD-10 has been confirmed by Output Check
1, insert the electrosurgical accessory into the endoscope and proceed with
the following steps.
The entire electrosurgical system cannot be tested when using PSD-10 in
combination with CD-3U/L or CD-4U/L. Be sure to perform the Output
Check 1 for the PSD-10 to ensure the proper connection of the system.
Connect the A-cord to the PSD-10 (ACTIVE connector) and to the electro-
surgical accessory (A-cord plug on SD-handle).
Only Olympus A-cord MA-255 is compatible with this equipment.
Push each end of the A-cord until a click is felt and heard.
Connect the smaller plug of the Check Lead to the PSD-10 TEST connector.
Connect the larger plug (spring-loaded clip) to the electrode tip/snare wire
which is protruding out of the endoscope.
Before connection, wipe the clip clean with gauze dampened with alcohol.
When attaching the clip to the wire, be careful not to deform or damage
the wire (snare, loop, etc.)
Depress the footswitch and ascertain that the TEST "OK" light (green) comes
on.
If the TEST "FAULT" light (red) comes on:
a. Connection between the A-cord and the electrosurgical accessory is de-
fective.
b. Wire is broken in the above line, or
c. The electrode tip is in contact with the metal portion of the endoscope
tip.
Do not touch the Patient-plate; burns may occur.
After performing this check, disconnect the Check Lead and the A-cord from
the PSD-10.