Olympus CKX-NU Manual de instrucciones - Página 27

Navegue en línea o descargue pdf Manual de instrucciones para Equipos de laboratorio Olympus CKX-NU. Olympus CKX-NU 36 páginas. Reflected fluorescence system

Olympus CKX-NU Manual de instrucciones
1
Fig. 15
3
2
1
Fig. 16
6
Attaching the Excitation Light Protective Shield
2
}The mounting holes for the excitation light protective shield
3
have a large size because they are also used with the
CKX-RCD condenser.
Insert the excitation light protective shield 1 from the bottom
of the condenser 2, and then tighten the two clamping
screws 3.
7
Attaching the U-LH50HG Lamp Housing
Insert the collector 1 of the lamp housing into the illuminator,
rotate the lamp housing so that its top panel 2 becomes
4
horizontal, then tighten the two clamping screws 3 on the
illuminator using the Allen wrench.
As the lamp housing surface becomes very hot during
operation, ensure that there is ample free space around
the lamp housing, especially above and below.
5
If the lamp socket cord 4 or power cord 5 of the CKX41
comes in contact with the lamp housing or its surround-
ings, the cord could melt and cause electric shock. Be
sure to position these cords at safe distances from the
lamp housing.
REFLECTED FLUORESCENCE SYSTEM
FOR CKX41
(Fig. 15)
(Fig. 16)
24