When the eyepiece is not equipped with the eyepiece micrometer
1
Rotate the pointer brightness control knob c of the discussion
attachment in the clockwise direction to turn ON the pointer ( ).
2
Set the pointer color switching lever d to the Down position (G). At
this time, if the pointer cannot be observed in the field of view, move
the pointer to the center of the field of view using the pointer control
lever e .
3
Rotate the diopter adjustment ring f of the right eyepiece to bring
the pointer into focus.
4
Rotate the interpupillary distance adjustment ring g of the left
eyepiece sleeve to bring the pointer into focus.
5
Set the pointer color switching lever d to the Up position (R), and make
sure that the pointer is into focus or not. When it is not into focus,
perform
and
repeatedly at both positions of the pointer color
3
4
switching lever d to bring the both pointers, red and green, into focus.
6
Place the specimen and rotate the coarse/fine focusing knobs to bring
the specimen into focus.
Diopter adjustment of assistant observer
Perform the same operation as "Diopter adjustment of main observer".
If the eyepiece is not equipped with the eyepiece micrometer, turn
ON the pointer ( ) and bring the specimen into focus at the main
observer position.