FH ORTHOPEDICS be POD TYLOS Arthrodese-Nail Manual - Página 7
Navegue en línea o descargue pdf Manual para Equipamiento médico FH ORTHOPEDICS be POD TYLOS Arthrodese-Nail. FH ORTHOPEDICS be POD TYLOS Arthrodese-Nail 8 páginas.
EXTRACTION OF THE MATERIAL
The lateral screws are simply unscrewed once the approach has been retur-
ned to (ref. 257381).
The polyethylene plug protecting the head of the calcaneal screw is cleared,
making it possible to screw its extractor into place (ref. 254597). The extractor
is screwed in backwards, making it initially possible to connect it to the screw
and then, when it reaches its blocking point, to unscrew it in a second stage
(Fig. 20).
For the nail, once you have returned to the plantar approach, screw in the M6
connection screw (ref. 253251) on to which the nail extractor (ref. 254597)
and its counterweight (ref. 255440) can be connected.
Remark : Make sure to maintain a lateral screw in position so as not to push
it into the tibia.
TYLOS
254597
255440
Tylos
TM
253251
20 - Extraction of the TYLOS nail
7