OPERATION

Before attempting any serial communication, first ensure that
the ZMT Analyzers connected to the computer terminal or host
computer by serial link are functioning correctly as individual
instruments. This is achieved by connecting all analog inputs,
applying the input signals and checking that the digital display
reads appropriately.
Ensure that the serial data connections to ZMT Analyzers
have been made correctly with respect to the computer
terminal, or host computer, interface. If the above check
appears satisfactory, test the serial communication by sending
an appropriate message from the compcorrectly correctly uter
terminal or host computer to a transmitter and observe if it
replies; thus establishing communication. If communication is
not established, check that the computer terminal, or host
computer, interface is correctly set up and that the plug-in
links within each transmitter are positioned correctly
– see Section 5.
Check that the parameters programmed in the instrument's
Serial Data Communication Page
the computer terminal or host computer – see Section 7.
If communication is still not possible or is erratic, check that the
computer terminal, or host computer, interface has pull-up and
pull-down resistors connected as shown in Figs. 9.1
Note. If no reply is received from the instrument within
160ms, retransmit the command. If after five command re-
entries, a satisfactory reply has not been received, the
communication link has been broken and must be
rechecked.
12
are compatible with those of
Computer Terminal or Host Computer
Pull-down
Tx-
Resistor
Tx+
Rx-
Rx+
Pull-down
Resistor
Fig. 9.1 Pull-up and Pull-down Resistors

10 SPECIFICATION

As detailed in the Operating Instructions (IM/ZMT), with the
following additions:
EIA Communication
Standards
4-wire mode
Parity
Odd
Even
Block check character
Transmission line length 1200m max.
Transmission speeds
2400 baud
4800 baud
9600 baud
+5V
Pull-up
Resistor
'A'
'B'
+5V
Pull-up
Resistor
0V
'A'
'B'
0V
RS422, RS485
None
Programmable
Programmable on or off
1200 baud
Programmable