Blue Tooth Connectivity
This scale comes with Blue Tooth Connectivity. Located in the front of the
scale under the indicator you will find the MAC address for the Blue Tooth
Communication.
Calibration
This scale has been factory calibrated and does not require calibration prior to use.
Cleaning
Clean the unit using a gentle cleaning agent. Under no circumstances shall abrasive or acid
cleaners, white spirit, benzene or similar products be used. Such substances can damage the
high-quality surfaces.
Troubleshooting
What do I do if...
... weight reading does not appear in the display window?
o Check the batteries
o Check that AC adaptor is plugged into unit and outside power source
... 0.0 does not appear before the weighing operation?
o Press the ZERO button to turn on the scale - there must not be any load on the
scale
... ---- appears instead of 0.0 before weighing?
o Press the ZERO button - there must not be any load on the scale
... one display window is illuminated constantly or not at all?
o The relevant display window has a fault - contact your Medline sales
representative or call 1-800-MEDLINE
... the display window shows a "?"?
o Battery voltage is running low - change the batteries within the next few days
... bAtt appears in the display window?
o Replace batteries
... ou Ld appears in the display window?
o Maximum weight load has been exceeded – must use alternative scale with
heavier capacity