digiO2 ETH-102 Manual de instrucciones - Página 11

Navegue en línea o descargue pdf Manual de instrucciones para Termómetro digiO2 ETH-102. digiO2 ETH-102 14 páginas. Bt infrared ear thermometer

Guidance and manufacturer's declaration-electromagnetic emissions
The ETH-102 is intended for use in the electromagnetic environment specified below.
The customer or the user of the ETH-102 should assure that it is used in such an
environment.
Emission test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations
Not applicable
/flicker emissions
IEC 61000-3-3
19
Electromagnetic environment-guidance
The ETH-102 uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The ETH-102 is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Guidance and manufacturer's declaration-electromagnetic immunity
ETH-102
The
is intended for use in the electromagnetic environment specified below.
ETH-102
The customer or the user of the
should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Electrostatic discharge(ESD)
+ 6 kV contact
IEC 61000-4-2
+ 8 kV air
Electrical fast transient/burst
+ 2kV for power supply
IEC 61000-4-4
lines
+ 1kV for input/output
lines
Surge IEC 61000-4-5
+ 1kV line(s) to line(s)
+ 2kV line(s) to earth
Voltage Dips, short
<5% UT(>95% dip in
interruptions and voltage
UT) for 0,5 cycle
variations on power supply
40% UT(60% dip in
input lines IEC 61000-4-11
UT) for 5 cycles
70% UT(30% dip in
UT) for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
Power frequency(50/60 Hz)
3 A/m
magnetic field IEC 61000-4-8
NOTE
UT is the a.c. mains voltage prior to application of the test level.
Compliance level
Electromagnetic
environment-guidance
+ 6 kV contact
Floors should be wood, concrete or
+ 8 kV air
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%
Not applicable
Mains power quality should be that of a
Not applicable
typical commercial or hospital
environment.
Not applicable
Mains power quality should be that of a
Not applicable
typical commercial or hospital
environment.
Not applicable
Mains power quality should be that of a
typical commercial or hospital
Not applicable
environment. If the user of the
ETH-102
requires continued
Not applicable
operation during power mains
interruptions, it is recommended that the
ETH-102
Not applicable
be powered from an
uninterruptible power supply or a battery.
ETH-102
3 A/m
The
power frequency
magnetic fields should be at levels
characteristic of a typical location in a
typical commercial or hospital
environment.
20