DIGIOXI PO-930 Manual del usuario - Página 6
Navegue en línea o descargue pdf Manual del usuario para Monitor DIGIOXI PO-930. DIGIOXI PO-930 19 páginas. Pulse oximeter monitor
A
Mechanical Description
.
Size
Weight
Color
B
Power Requirements
.
AC voltage Input
Internal Battery
Battery Duration
Recharge Time
C
SpO
- Pulse Oximeter
.
2
SpO
Range
2
SpO
Accuracy
2
Pulse Rate Range
Pulse Rate Accuracy
Pulse Tone
Sensor Types
D
Heart Rate
.
Heart Rate Range
Heart Rate Accuracy
E
Display
.
Type
Dot Matrix
Effective Display Area
Viewing Angle
Operators Manual Pulse Oximeter Monitor DIGIOXI "PO-930"
SECTION 2 - Technical specification
: 94x254x254mm
: 3,5 Kg
: White and Black
: 120VAC; 60Hz (220VAC and/or 50Hz Optional)
: 12 VDC/2,9Ah Sealed
: 7 Hours.
: 18 Hours in use on AC line.
: 0% to 100% adult/pediatric/neonate
: ± 2% (70 - 100%), ± 3% (50 - 69%) unspecified (0 - 49%)
: 30 to 250 BPM
: ± 2%
: Pitch adjusted by SpO
: Finger, Universal, Wrap, Disposable.
: 30 to 255 BPM
: 2%
: High Intensity CFL-LCD
: 240x64 Dots
: 178 mm x 69 mm x 25mm
: 150°
Digicare Biomedical Technology Inc.
2
4