Siemens ULTRAMAT 6F Manual de instrucciones - Página 14
Navegue en línea o descargue pdf Manual de instrucciones para Equipamiento industrial Siemens ULTRAMAT 6F. Siemens ULTRAMAT 6F 49 páginas. Gas analyzer for the determination of ir-absorbing gases
También para Siemens ULTRAMAT 6F: Manual de equipamiento adicional (20 páginas), Manual de seguridad (24 páginas)
3.2
Installation location requirements
CAUTION
Direct sunlight
Device damage.
The device can overheat or materials become brittle due to UV exposure.
Protect the device from direct sunlight.
•
Make sure that the maximum permissible ambient temperature is not exceeded. Refer to the information in Technical
•
specifications (Page 37).
WARNING
Insufficient air circulation
Risk of fire.
The device may overheat or start burning in case of insufficient air circulation.
Ensure sufficient air circulation in the room. Refer to the information in Technical specifications (Page 37).
•
NOTICE
Incorrect mounting
The device can be damaged, destroyed, or its functionality impaired through improper mounting.
Before installing ensure there is no visible damage to the device.
•
Make sure that process connectors are clean, and suitable gaskets and glands are used.
•
Mount the device using suitable tools. Refer to the information in Technical specifications (Page 37).
•
3.2.1
Strong vibrations
CAUTION
Strong vibrations
Vibrations can influence the result of a measurement.
The analyzer must therefore only be installed at a location which is free of vibration.
•
Ensure that there is little vibration when using the device. The maximum vibrations must not exceed 5 m/s².
•
14
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
A5E45779389002-01, 09/2018