Siemens SITRANS L Series Manual de instrucciones - Página 15

Navegue en línea o descargue pdf Manual de instrucciones para Interruptor Siemens SITRANS L Series. Siemens SITRANS L Series 33 páginas. Capacitance transmitters
También para Siemens SITRANS L Series: Instrucciones de uso Compact (45 páginas), Manual de instrucciones (50 páginas), Manual de instrucciones (44 páginas)

Siemens SITRANS L Series Manual de instrucciones
Note
• Before opening the lid, ensure there are no dust deposits around the device. Ensure that
the atmosphere around the device is settled.
• Use a fuse for the signal output and power supply (max. 10 A, fast or slow, HBC, 250 V).
• Make sure the main voltage does not exceed the maximum voltage listed on the product
label and the setting of the voltage selector.
• In case of inexpert handling or handling malpractice, the electric safety of the device
cannot be guaranteed.
• Provide protection for relay contacts to protect the device against spikes if inductive loads
are connected.
6.1

Universal voltage, DPDT, fail-safe

Power supply:
24 V DC ± 15 %
50/60 Hz ± 10 %
max. 10A, fast or slow, HBC, 250 V
Signal and alarm output:
Relay DPDT contact
max. 250 V AC, 5A non-inductive
max. 30 V DC, 4A, non-inductive
External fuse: max. 10A, fast or
slow, HBC, 250 V
Including ± 10 % of EN 61010-1.
1)
With fail-safe option, contact is open when de-energized.
2)
SITRANS LPS200
Operating Instructions, 04/2019, A5E34210883-AB
max. 4 W 22 ... 230 V
1)
max. 10 VA External fuse:
1)
Max. 4 mm
(AWG12)
2
PE
Fail-safe
2)
Signal output DPDT
15