Ambient conditions
Ambient temperature on the housing
Storage and transport temperature
Process conditions
Measured variable
Process pressure
Maximum allowable operating pressure
Process temperature (thread or flange
temperature)
Ʋ SITRANS LVL200E of 316L/Alloy C22
(2.4602)
Process temperature (thread or flange
temperature) with temperature adapter
(option)
Ʋ SITRANS LVL200E of 316L/Alloy C22
(2.4602)
Ʋ SITRANS LVL200E enamelled
Ʋ SITRANS LVL200E with ECTFE
coating
Ʋ SITRANS LVL200E with PFA coating
-50 °C
(-58 °F)
Fig. 25: Ambient temperature - Process temperature
1
Process temperature in °C (°F)
2
Ambient temperature in °C (°F)
3
Temperature range with temperature adapter
*7ML19985KY01*
-40 ... +70 °C (-40 ... +158 °F)
-40 ... +80 °C (-40 ... +176 °F)
Limit level of liquids
-1 ... 64 bar/-100 ... 6400 kPa (-14.5 ... 928 psig)
The process pressure is dependent on the process
fitting, for example Clam p or flange (see the following
diagrams)
100 bar/10000 kPa (1450 psig) or 1.5 times process
pressure
The function of the instrument is ensured up to an
operating pressure of 100 bar/10000 kPa (1450 psig)
at a maximum process temperature of +50 °C (+122 °F)
(only with threaded versions).
-50 ... +150 °C (-58 ... +302 °F)
-50 ... +250 °C (-58 ... +482 °F)
-50 ... +200 °C (-58 ... +392 °F)
-50 ... +150 °C (-58 ... +302 °F)
-50 ... +250 °C (-58 ... +482 °F)
2
70 °C
(158 °F)
40 °C
(104 °F)
0 °C
(32 °F)
50 °C
(122 °F)
-40 °C
(-40 °F)
SITRANS LVL200E - Operating Instructions
3
100 °C
150 °C
200 °C
(212 °F)
(302 °F)
(392 °F)
1
250 °C
(482 °F)
31