Biuged BGD 527 Manual de instrucciones - Página 7

Navegue en línea o descargue pdf Manual de instrucciones para Equipos de prueba Biuged BGD 527. Biuged BGD 527 8 páginas. Wet abrasion scrub tester

Biuged BGD 527 Manual de instrucciones
6.0 Report
Record the number of cycles to remove one continuous thin line of paint film across the 12.7mm
width of the shim. Stop the machine and wipe off the shim area to determine, if necessary, if the end
point is achieved.
Test two of the drawdowns. If the two results differ by more than 30% of the average, test the third
drawdown. If one of the three results is obviously discrepant, discard it, then average the results from the
remaining drawdowns.

7.0 Water Container

7.1 Washing Liquid can't be used cyclically in case the water pipe is blocked. The operator should
prepare another container to receive the waste washing liquid which is discharged from drain pipe.
7.2 Insert the feed-water pipe into the quick fittings of water container and ensure it's tight
7.3 Connect the plug of water pump to socket of pump which is located the side of machine. Connection
for water container is finished.
7.4 Please clean the water container after every test, the step is as below:press the blue part of quick
fittings and pull out the soft water pipe at the same time. See the below picture.
7.5 Open the cover of water container then clean it

8.0 Maintenance

8.1 Clean and machine with clean and soft cloth after using, keep it in the dry and clean environment。
8.2 Replace the brush when the bristles have become worn to a point that the bristles extend less than
15mm from the block. Mark the ends of the brush to ensure it is oriented in the same direction for each
use.
8.3 Don't cut off the supply of washing liquid when pump is working.