AEMC instruments 100.185 Manual del usuario - Página 3

Navegue en línea o descargue pdf Manual del usuario para Instrumentos de medida AEMC instruments 100.185. AEMC instruments 100.185 4 páginas. Ac current probe

AEMC instruments 100.185 Manual del usuario
Phase Shift:
≤ 3.5° from 1 to 100A, 50 to 60Hz
(*Reference conditions: 23°C ± 3°K, 20 to
85% RH, external magnetic field < 40A/m,
no DC component, no external current
carrying conductor, test sample centered.)
Load impedance 1Ω.
Overload:
170A continuously
Frequency Range:
30Hz to 10kHz
Load Impedance:
5Ω max
Working Voltage:
250V AC
Common Mode Voltage:
30V AC
MECHANICAL SPECIFICATIONS
Operating Temperature:
-13° to 122°F (-25° to 50°C)
Storage Temperature:
-40° to 176°F (-40° to 80°C)
Temperature Influence:
< 0.2% per 10° K
Maximum Conductor Size:
0.47" Ø max. (12mm)
Dimensions:
1.26 x 4.53 x 0.87"
(32 x 115 x 22mm)
Weight: 5.6 oz (160g)
Colors:
Dark gray handles with red cover
Polycarbonate Material:
Handle: 10% Fiberglass charged
polycarbonate UL 94 V0
0.561.8187
www.
.com
Output:
Two standard safety banana jacks (4mm)
SAFETY SPECIFICATIONS
Electrical:
3kV 50/60 Hz dielectric
30V max common mode between output
and ground
ORDERING INFORMATION
AC Current Probe MN185 ... Cat #100.185
Accessories:
Lead - Set of 2, color-coded (red/black)
5 ft Safety Leads
(1000V CAT IV) ................... Cat #2152.24
Adapter - 4 mm non-insulated for
Safety Leads....................... Cat #1017.45
information@itm.