SET-UP
Prior to use
Before use, the Gas-Pro should always be checked for any signs of physical
damage.
i
Read and understand the User and Operator manual prior to use.
Gas-Pro orientation
Figure 1: Gas-Pro
À
Á
È
Æ
Å
Â
Ç
Ä
Ã
À D-ring
à Operator button Æ +ve Safety™ indicator
Á Alarm bars Ä Sensor apertures Ç Dual colour LCD display
 Sounder
Å Pump inlet/outlet È Alligator clip
DISPLAY
Home Screen
Gas-Pro simultaneously displays up
to 5 gases.
The dual colour backlight aids
recognition of alarm events.
The home icon will invert to show normal operation.
The backlight is green and changes to red during an alarm highlighting
the gas in alarm.
2
OPERATION
Switching On/Off
In clean air turn on the Gas-Pro by holding down the operator button
for 3 audible blips. The Gas-Pro will warm up, going through a series of
automatic processes.
To turn off, press and hold the operator button. During the 4 second
countdown hold the button down until the countdown has finished.
Quick View
A momentary press of the operator button allows the user to see the gas
type and unit, battery level, user and +ve Safety™ status.
Alarms
Gas-Pro offers audible, visual and vibrate alarms as standard.
Each gas has 2 instantaneous alarms and toxics also have TWA alarms.
When in alarm the display backlight will change from green to red and
É
highlight the gas in alarm.
When the Gas-Pro warns of a low battery, the sounder will emit an audible
double blip every 5 seconds and the +ve Safety™ LED will go red. This
means the battery has at least 30 minutes of battery life remaining. After 30
É Charging cable
minutes the Gas-Pro will enter full alarm state and the battery icon will flash
Certification label
empty as shut down is imminent.
Backlight
The backlight is configurable but set as standard to come on at the press
of the operator button or during alarm.
Pump
If the internal pump option has been chosen, the pump is activated by
attaching the pumped flow plate.
Ensure the gasket is clean and undamaged prior to placement.
The pump test will activate every time the flow plate is attached the user
will be prompted to cover the inlet to ensure a tight seal. Gas-Pro will
indicate a pass or fail, if failed the flow plate should be removed, checked
and reapplied.
USER MENU
the Portables-Pro PC application.
an internal pump, allows authorised users to turn the pump on or off.
sounder volume between 2 levels.
3
Access the user menu by double clicking the operator button, then
scroll through the menu using single clicks. Once the correct function
is highlighted, double click to choose.
Zero
During start-up the operator has the option to perform an autozero, or
leave the countdown if an autozero is not required/possible.
A manual zero can be performed at any time following start up and is
needed less than 15 minutes prior to running a manual calibration.
Time Weighted Average (TWA)
This function allows the 8 hour TWA to be reviewed.
Pre-Entry Check (PEC)
This function is intended for sampling air of unknown quality before
gaining access to it (e.g. going under ground through a manhole
cover) thereby avoiding unnecessary exposure.
Peak Review
This function allows the highest level of each gas detected to be
reviewed during the last 8 hours, 12 hours or since turned on.
Settings
User setting: Up to 5 different users can be loaded into Gas-Pro using
Pump setting: This function, which is only present if the Gas-Pro has
Sounder volume: This function allows authorised users to change the
4