Influence of Conductor in Jaw Opening:
0.1 % of Reading
Influence of Frequency:
From (30 to 48) Hz: < 0.5 % of R
From (65 to 1000) Hz: < 1 % of R
From (1 to 5) kHz: < 2 % of R
MECHANICAL SPECIFICATIONS
Operating Temperature:
(14 to 122) °F (-10 to 50) °C)
Storage Temperature:
(-4 to 158) °F (-20 to 70) °C
Influence of Temperature:
< 0.15 % per 10 °K
Influence of Humidity:
From (10 to 90) % RH: 0.1 %
Jaw Opening:
2.25 in (57mm) max
Maximum Conductor Size:
2.05 in (52 mm)
Envelope Protection: IP 40 (IEC 529)
Drop Test: 1 m (IEC 68-2-32)
Mechanical Shock: 100 g (IEC 68-2-27)
Vibration:
(5 to 15) Hz, 0.15 mm (IEC 68-2-6)
(15 to 25) Hz, 1 mm
(25 to 55) Hz, 0.25 mm
Material:
Handles and jaws: Polycarbonate UL94V0
Dimensions:
(4.37 x 8.50 x 1.77) in (111 x 216 x 45) mm
OPERATION
Please make sure that you have already read and fully understand the WARNING section on
page 1.
Making Measurements with the AC Current Probe Models SR701/SR740
Connect the black lead of the current probe to "common" and the red lead to the AC
current input on your DMM or other current measuring instrument. Select the appropriate
current range (2 A
arrow pointed toward the load. If the reading is less than 200 mA, select the lower range
until you obtain the best resolution. Read the value display on the DMM and multiply it
by the probe ratio (1000/1). (If reading = 0.659 A, the current flowing through the probe is
0.659 A x 1000 = 659 A
For best accuracy, avoid if possible, the proximity of other conductors which may create
noise.
range). Clamp the probe around the conductor to be tested with the
AC
).
AC
Weight:
1.21 lbs (550 g)
Output:
SR701: (2) standard safety banana jacks (4
mm)
SR704: 5 ft (1.5 m) lead with safety 4 mm
banana plug
SAFETY SPECIFICATIONS
Electrical:
Double insulation or reinforced insulation be-
tween the primary or secondary and the outer
case of the handle conforms to
IEC 1010-2-032.
Common Mode Voltage:
600 V CAT III, Pollution Degree 2
Electromagnetic Compatibility:
Emission and immunity in an industrial envi-
ronment according to EN61326-1
ORDERING INFORMATION
AC Current Probe SR701 ........Cat #2116.29
AC Current Probe SR704 ........Cat #2116.30
Accessories:
Lead, set of 2, 5 ft Safety Leads
(1000 V CAT IV) ......................... Cat #2152.24
Adapter BNC (Male) - Banana (Female)
(XM-BB) (600 V CAT III) ............. Cat #2118.46
Banana plug adapter
(to non-recessed plug)................ Cat #1017.45