aeolus UC-1901BK Manual del usuario - Página 4
Navegue en línea o descargue pdf Manual del usuario para Producto para el cuidado de mascotas aeolus UC-1901BK. aeolus UC-1901BK 10 páginas. Professional multifunctional veterinary intensive care unit
Choosing The Working
Space For The Unit
Care should be taken when choosing the working space for this unit.
The unit should be used indoors only, in a cool
dry room with stable temperature and humidi-
ty, kept out of direct sunlight, away from any
heat sources, excess moisture or naked
flames.
This unit should only be operated in a
temperature range of 10° to 40° C (50° to 103°
F). It should not be placed in the direct path of
a room air conditioning outlet.
The full unit weighs 300 kg (660lbs); it should
be placed on a level stable floor.
It should never be placed on a soft or uneven
surface like bedding materials or shaggy
carpets.
The unit should be placed in a well-ventilated
position, and all inlets and outlets should be
unobstructed, there should be a 15cm or
6-inch clearance on all sides.
≥15cm
04
Parameters
Professional Veterinary
Product Name
Intensive Care Unit
Model Number
UC-1901M/L/BK
1509mmW×884mmD×1899mmH
Dimension
(59.5"W×35"D ×75"H)
Approximate 300kg (660lbs)
Weight
AC100~120V /220~240V
Power Supply
50/60Hz (Grounded)
4 Compartments Controlled
Separately, Cooling and Heating.
Two Medium Compartments on
Environment
Top Tier, One Large Compartment
Control
on Bottom Tier which could be
divided into two medium com-
partments when needed.
Ultra Violet Sterilization, High
Air Purification
Concentration Negative Ion Air
Purification.
Working
-10~40°C (Indoor)
Condition
Temperature
15~38°C, Tolerance at 0.5°C
Setting Range
Oxygen Concen-
21~65%, Tolerance at ±1%.
tration Setting
Carbon Dioxide
200 -5000PPM,
Monitoring and
Tolerance at ±10PPM,
Removal
Built-in CO2 Sensor and Remover
O2 Concentration Alarm, Tempe-
rature Alarm, Sensor Alarm, Door
Opening Alarm, CO
2
Alarms
Concentration.
Alarm, Emergency Ventilation Port
Malfunction Alarms.
Automatic Dehumidification,
High/Middle/Low Three Stages,
Dehumidification
High: 45%~50%,
Middle: 40%~44%,
Low: 30%~39%.
Control Panel
8.4", 171mmW×128mmH
Screen Type
Resistive
Max. Power
1.3KW
Consumption
Average Power
0.5KW
Consumption
Power Supply
Emergency Ventilation Port
Failure Protection
Negative Ion
7.2×106pcs/cm3
Generator
Negative Ion Generator
Soda Lime CO2 Absorption
Accessories
Powder, Equipment Mount Grid,
Door Shade Curtain.
Built-in Printer
Smart Control Needle Type Printer
Ultra-violet
253nm UV Sterilization
Sterilization
Blue Light
415nm Blue Therapeutic Light
Therapy
Treatment
Medical
≥0.2ml/min, particle size:
Nebulizer
0.5 -2μm, noise≤ 40dB(A)
Pure White Cold Light for
Examination, 10 Stage Adjustable
Yellow .
LED Illumination
Comfort Light for Illumination.
Low/Middle/High, Three-Stage Speed
Control.
Defaults:
Oxygen Input On: Low Speed.
Ventilation Fan
Temperature Control: High Speed
Control
Emergency Port Open: High Speed
Ultra-violet Sterilization On: Middle
CO2 Concentration Level Alarm On:
High
05