g) Foam on the level absorbs the acoustic wave re-
flection which might cause malfunction of the level
meter. For mounting find the location where the
foaming is as low as possible (Fig. 5).
h) Emitted acoustic signal must not be affected by near
objects (ladders, mixers, propellers, etc.), stream
of filling, air flow, etc. (Fig. 6).
Fig. 6: Minimum distance of the level meter to nearby objects
i) Do not install the level meter in or above the filling point (Fig. 7).
Fig. 7: Level meter installation outside the influence of filling circulation
j) In case the level of bulk-solid materials is measured, the measurement range is reduced. We
recommend to consult the use with the manufacturer.
8
in the tank
!
Fig. 5: Thick foam on the surface
x > 1/12 c
ULM–53L–02 ; 10
(min. 200 mm)
x > 1/8 c
ULM–53L–06
(min. 200 mm)
x > 1/10 c
ULM–53L–20
(min. 200 mm)
x – Distance to the edge of the longest object
c – Measurement range of the level meter
ULM–53L
© Dinel, s.r.o.