CALIBRATION procedure

The calibration is semi-automatically because the program sets and suggests the execution
sequence, but is the operator that must perform properly the sequence. At the end of the
sequence the program scan the data and calculates the average, the errors, the uncertainty
of the measurements.
"Calibration Characteristics " setting
 Points of measurement: set the number of points of measurement (eg "5" maximum
"10").
 Cycles of measurement: set the number of cycles of measurement (eg "2" maximum
"3").
 Type of measure: Set "Compression" if the calibration is performed with increasing
positive set "Tensile" if the calibration is performed with increasing negative; calibration
is possible only in tension with load cells bidirectional.
 Interpolate Measured Values : This selection allows you
to interpolate the measurements performed on the
instrument to be calibrated to the actual load that has been created. This situation is
typical in testing machine where it is very difficult to achieve precise nominal load.