Alcolizer Druglizer LE5 Manual del usuario - Página 4
Navegue en línea o descargue pdf Manual del usuario para Equipos de prueba Alcolizer Druglizer LE5. Alcolizer Druglizer LE5 16 páginas. Charging cradle
También para Alcolizer Druglizer LE5: Manual del usuario (20 páginas), Manual del usuario (19 páginas), Instrucciones de montaje (3 páginas)
conjugate pad where drugs in the sample react with tagged antibody specific for that drug.
The sample continues to wick by capillary forces along the nitrocellulose membrane, onto
which is attached the same drug of interest. If no drug is present in the oral fluid sample the
tagged antibody can bind to the drug bound to the nitrocellulose and a red line is produced.
If drugs are present in the oral fluid sample at concentrations greater than the cut-off
concentration, the tagged antibody sites will be saturated with the drug and will not be able
to bind to the nitrocellulose bound drug, hence no result is produced. As the drug
concentration in the sample increases the colour intensity of the test line decreases. The test
strip is not intended to be viewed visually rather it is intended to be interrogated by a meter.
The instrument uses a sophisticated algorithm the determine the result.
3 MATERIALS PROVIDED
The Druglizer LE5 is supplied with the following:
• Druglizer LE5 main module.
• Druglizer LE5 drug test module.
• USB cable.
• Battery Chargers (240VAC and 12VDC vehicle).
• Silicon protective cover, cartridge plug and wrist strap.
• Soft shoulder bag or hard plastic carry case (optional).
• Bluetooth printer (optional).
Updated user Manual and Alcoconnect can be downloaded from the
https://www.alcolizer.com/software-updates/
Cartridges are not included with the purchase of an instrument. Cartridges are sold separately
in cartons of 25 tests.
4 WARNINGS AND PRECAUTIONS
The Druglizer LE5 has been manufactured and tested according to accepted technical
principles. However, improper use may still lead to measuring errors and misinterpretations
and the precautions should be observed:
• For professional use only.
• Always store instrument with Cartridge plug when not in use.
• Always transport the instrument in the carry case or bag.
• Always check the battery levels before use and recharge the instrument if necessary.
• Avoid storage Druglizer LE5 main module at low temperatures (< 0°C).
REF: Druglizer LE5 Drug Tester User Manual V1
Page 4