8 Calibration

8.1 Calibration Gas Value Input

Input the reference gas calibration value into Calibration Gas menu as follows:
 Press the Test key to open the Calibration Gas menu input function (refer Figure 1).
 Press the Test key again to start input function. The calibration gas first digit value
(underlined) can be changed by pressing either of the Up/down keys to required first
digit number value (refer Figure 2).
 Press the Test key to accept the entered value and to proceed to the next digit
number input value.
 Repeat the above procedure to set and accept each digit's calibration gas value to
the required (currently used) Wet or Dry gas calibration reference standard e.g.
0.100g/100mL (refer Figure 3).
REF: LE5 Calibration Instructions_Whole Units V5
Figure 1 - Calibration Gas
Figure 2 - Setting Calibration Gas Value
Page 6
© Alcolizer Technology