8 Calibration
8.1 Calibration Gas Value Input
Input the reference gas calibration value into Calibration Gas menu as follows:
Press the Test key to open the Calibration Gas menu input function (refer Figure 1).
Press the Test key again to start input function. The calibration gas first digit value
(underlined) can be changed by pressing either of the Up/down keys to required first
digit number value (refer Figure 2).
Press the Test key to accept the entered value and to proceed to the next digit
number input value.
Repeat the above procedure to set and accept each digit's calibration gas value to
the required (currently used) Wet or Dry gas calibration reference standard e.g.
0.100g/100mL (refer Figure 3).
REF: LE5 Calibration Instructions_Whole Units V5
Figure 1 - Calibration Gas
Figure 2 - Setting Calibration Gas Value
Page 6
© Alcolizer Technology