Alcolizer Druglizer LE5 Manual del usuario - Página 12

Navegue en línea o descargue pdf Manual del usuario para Unidad de control Alcolizer Druglizer LE5. Alcolizer Druglizer LE5 20 páginas. Charging cradle
También para Alcolizer Druglizer LE5: Manual del usuario (16 páginas), Manual del usuario (19 páginas), Instrucciones de montaje (3 páginas)

Alcolizer Druglizer LE5 Manual del usuario
Perform a Mouthpiece Test as follows:
• Press the Test key, the instrument will perform a background self-test while Please
wait....is displayed on the screen.
If the self-check is unsuccessful an error code will be displayed, the instrument will not accept a
breath sample. Contact an Alcolizer representative if an error code is displayed. If error code 81 is
• BLOW will be displayed when the instrument is ready (Refer Figure 7).
• After taking a deep breath, the test subject places their mouth over the mouthpiece
inlet (ribbed end) and blows a gentle to medium breath sample into the mouthpiece
(Refer Figure 8).
• The exhaled breath volume is indicated via a sample progress bar; scrolling left to
right at the bottom of the screen and a graph showing air flow over time in seconds
(Refer Figure 9).
REF: Alcolizer LE5 Alcohol Tester User Manual V15 docx
present: follow onscreen prompt to restart the device.
Figure 7- Blow is displayed when the instrument is ready
Figure 8- Providing a Sample
Note
Page 12
© Alcolizer Technology