AlcoSense Precision + Manual del usuario - Página 5

Navegue en línea o descargue pdf Manual del usuario para Equipos de prueba AlcoSense Precision +. AlcoSense Precision + 10 páginas.

condItIonS For tEStIng

• Use a clean mouthpiece for each test.
• Do not allow samples containing smoke, tobacco or other
chemical products to enter the unit, as damage to the fuel
cell can occur.
• There must be an interval of at least 15 minutes after alcohol
consumption as alcohol residue in the mouth may give a false
reading.
• Such residue may also be left by aromatic drinks,
mouthwash, medicines and drops, as well as by belching
and vomiting. Rinsing your mouth with water or non-alcoholic
drinks does not substitute for an interval.
• A person's blood alcohol concentration (BAC) can continue
to rise for up to 2 hours after drinking alcohol. If a caution
results or a result is close to the designated BAC, it is
recommended to wait and retest after 20 minutes. It can take
10 hours or more for a person's BAC to return to zero after a
high BAC has been reached.
• A further test should be carried out later in the day or the
following morning.
• working temperature range: 5°c to 40°c.
Storage temperature: -15°c to 65°c.
lIAbIlIty For propEr FunctIon or dAmAgE
The liability for proper functioning of the instrument is
irrevocably transferred to the owner or operator in the
event the instrument is serviced or repaired by personnel
not employed or authorised by Andatech. In such an
event, the warranty will be void.
5