Direct Healthcare Group Dyna-Form Mercury Advance Manual del usuario - Página 12
Navegue en línea o descargue pdf Manual del usuario para Equipamiento médico Direct Healthcare Group Dyna-Form Mercury Advance. Direct Healthcare Group Dyna-Form Mercury Advance 16 páginas.
También para Direct Healthcare Group Dyna-Form Mercury Advance: Manual del usuario (9 páginas)
9. Technical Specification
Declaration – electromagnetic immunity – for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The MAT1260001 is intended for use in the electromagnetic environment specified below.
The customer or the user of the system should ensure that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000- 4-6
Radiated RF
IEC 61000-4-3
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the Span system is used exceeds the applicable RF compliance level above, the system should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the system.
b. Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
1 2
D Y N A - F O R M M E R C U R Y A D V A N C E T R O L L E Y
IEC 60601 test level
Compliance level
3 Vr
3 Vr
ms
150 kHz to 80 MHz
3 V/m
10 V/m
(Professional
Healthcare Environment)
10 V/m
(Home
Healthcare Environment)
80 MHz at 2.7 GHz
D I R E C T H E A L T H C A R E G R O U P . C O M
Electromagnetic environment – guidance
Portable and mobile RF communications equipment
ms
should be used no closer to any part of the CT515,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter
Recommended separation distance
d = 1.167√P
d = 1.167√P
d = 2.333√P
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
be less than the compliance level in each frequency
range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
80 MHz to 800 MHz
800 MHz to 2.5 GHz
b
should
a