Calex PyroCAN Series Manual del usuario - Página 8

Navegue en línea o descargue pdf Manual del usuario para Accesorios Calex PyroCAN Series. Calex PyroCAN Series 8 páginas.

OPERATION

Once the sensor is in position and the appropriate power, air, water, and cable connections are secure,
the system is ready for continuous operation by completing the following simple steps:
1. Turn on the power supply
2. Turn on the CAN instrumentation
3. Read / monitor the temperature
IMPORTANT
Be aware of the following when using the sensor:
• If the sensor is exposed to significant changes in ambient temperature (hot to cold, or cold to hot),
allow 20 minutes for the temperature to stabilise before taking or recording measurements.
• Do not operate the sensor near large electromagnetic fields (e.g. around arc welders or induction
heaters).
Electromagnetic interference can cause measurement errors.
• Wire must be connected only to the appropriate terminals.

MAINTENANCE

Our customer service representatives are available for application assistance, calibration, repair, and
solutions to specific problems. Contact our Service Department before returning any equipment. In
many cases, problems can be solved over the telephone. If the sensor is not performing as it should,
try to match the symptom below to the problem. If the table does not help, call Calex for further advice.

Troubleshooting

Symptom
No output
Erroneous temperature
Erroneous temperature
Erroneous temperature
LENS CLEANING
Keep the lens clean at all times. Any foreign matter on the lens would affect measurement accuracy.
Blow off loose particles (if not using the air purge accessory) with an air 'puffer'.
GUARANTEE
Calex guarantees each instrument it manufactures to be free from defect in material and workmanship
under normal use and service for the period of one year from the date of purchase. This guarantee
extends only to the original buyer according to Calex terms and conditions of Sale.
Probable Cause
No power to sensor
Incorrect wire connection
Faulty sensor cable
Field of view obstruction
Ihr Schweizer Industriepartner
Solution
Check power supply
Check wire colour codes
Verify cable continuity
Remove obstruction
www.digiparts.ch
Issue A - March 2017