También para Direct Supply Panacea: Manual del usuario (16 páginas), Manual del usuario (5 páginas), Manual del usuario (16 páginas), Manual del usuario (20 páginas), Manual del usuario (2 páginas), Manual del usuario (5 páginas), Manual del usuario (2 páginas), Manual del usuario (2 páginas), Manual del usuario (2 páginas), Manual del usuario (2 páginas), Manual del usuario (2 páginas)

Transportation & CPR
Transportation
1. Disconnect the hose from the pump and quickly place the cap over the end of the hose to prevent
air loss.
2. Check the fit of the cap for leaks.
3. Verify the red CPR valve at head-end of the mattress is closed and no air is escaping.
4. Remove pump from bed/footboard.
You may now transport the resident while in bed. A properly sealed mattress will continue to support
a resident for approximately one hour without power. It is not possible to provide an exact time
because it is dependent upon on the resident's weight. Reattach the pump as soon as possible to
restore mattress performance.
Prior to transporting the pump unit, turn the Power switch to OFF and unplug the power supply.
• WARNING: DO NOT attempt to use the mattress as a carrying device, as it is not
designed to carry a resident's weight unsupported.
CPR
The standards for life support by the American Heart Association recommend a hard, level surface for
performing CPR, moving the person to the floor if possible. If unable to move the resident to another
surface to perform CPR, place a crash board underneath the resident and follow appropriate CPR
procedures.
To deflate the mattress for CPR function, disconnect the hoses from the control unit and remove
the CPR plug from the hose located at the head end of the mattress. To remove the CPR plug,
squeeze the red plug cover labeled "CPR" and pull gently. The control unit should be turned off and/or
removed to prevent the mattress from re-inflating during CPR.
In case of emergency, use a scissors or knife to cut the hose and/or the air cushions in the mattress.