También para Direct Supply Panacea: Manual del usuario (16 páginas), Manual del usuario (5 páginas), Manual del usuario (16 páginas), Manual del usuario (20 páginas), Manual del usuario (2 páginas), Manual del usuario (5 páginas), Manual del usuario (2 páginas), Manual del usuario (2 páginas), Manual del usuario (2 páginas), Manual del usuario (2 páginas), Manual del usuario (2 páginas)

EMI (Electromagnetic Interference)
This product conforms to the electromagnetic compatibility requirement of IEC 60601-1-2:2014
standard. This product shall be used and operated in accordance with the instructions and
Manufacturer's Statement below.
Manufacturer's EMC Declaration
• WARNING: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the device.
Otherwise, degradation of the performance of this equipment could result.
• WARNING: Do not use this product next to or stacked with other devices.
The monitor is suitable for use in the specified electromagnetic environment. The customer and or user
of the monitor should assure that it is used in an electromagnetic environment as described below:
1.1 Compliance Information for Emissions Test
EMISSIONS Test
Radiated Emissions CLSPR 11
Radiated Emissions CLSPR 11
Harmonic Distortion IEC 61000-3-2
Voltage Fluctuations and Flicker IEC 6100-3-3
Electromagnetic Immunity (IEC60601-1-2)
1.2 Compliance Information for Immunity Test
IMMUNITY Test
Electrostatic Discharge (ESD)
IEC 61000-4-2
Power Frequency Magnetic Fields
IEC 61000-4-8
Electrical Fast Transients
IEC 61000-4-4
Surges IEC 61000-4-5
Voltage Dips and Interruptions
IEC 61000-4-11
Line to Line ±1KV Surges
IEC 61000-4-5
Conducted Disturbances Induced by RF Fields
IEC 61000-4-6
Compliance Level
±8 KV contact
±15 KV air
30 A/m
50 Hz
±2 KV for power supply lines
Line to Line ±1KV
Dips: 100% During 0.5 Periods, at 0°, 45°, 90°,
135°, 180°, 225°, 270°, 315°
100% During 1 Periods, at 0°
30% During 25/30 Periods, at 0°
Interruptions: 100% During 250/300 Cycles
10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
3V
0.15 MHz - 80 MHz
6 V in ISM Bands Between 0.15 MHz and 80 MHz
80% AM at 1 KHz
1-800-634-7328
Compliance
Group 1
Class B
Class A
Complies