Nikon Eclipse E400 Manual de instrucciones - Página 16
Navegue en línea o descargue pdf Manual de instrucciones para Microscopio Nikon Eclipse E400. Nikon Eclipse E400 21 páginas. Teaching head/multi teaching head
También para Nikon Eclipse E400: Manual de instrucciones (33 páginas), Manual de instrucciones (33 páginas)
5
Assembly for the multi-viewing model by 10 persons
1 Check that the serial numbers on the two components (on the main section and the
relay section) of the MTH relay unit are identical. (If the numbers are different,
observation will be adversely affected. In this case, please contact your dealer or
nearest Nikon representative.) Two sets of MTH relay units are required. Assemble
those components with identical serial numbers to each other.
2 Check that the serial numbers on the two components (on the main observer's section
and the secondary observer's section) of the teaching head S-S are identical. (If the
numbers are different, observation will be adversely affected. In this case, please
contact your dealer or nearest Nikon representative.) Four sets of teaching head S-S are
required. Assemble those components with identical serial numbers to each other.
3 Adjust the height of the support stands so that the main observer section and the
secondary observer sections are horizontal. This step is facilitated by first adjusting the
support stands to an approximate height before attaching the stands.
7
ND filter
(φ28 mm)
5
Assemble in the
same manner
as that for the
right side.
Clamp for 5
4
Clamp for 4
Check that the
serial numbers
are the same.
8
13
Clamp
for !3
6
Clamp for !0
Place
horizontally
10
Clamp
for !1
MTH relay unit
1
Main section
2
III. Assembly
3
Check that the
Nikon
Y-THF
serial numbers of
JAPAN
0100001
the teaching head
S-S are the same.
15
Clamp
for !5
12
Place
horizontally
17
Clamp for !2
11
9
Height adjustment
clamp screw
Adjust to an
approximate height
and readjust later.
Y-THF
Nikon
0100001
JAPAN
Clamp for !7
16
14
15