Nikon N-STORM Manual de instrucciones
No.
Name
12
[Graph] checkbox
13
[Auto LP] button
14
[Max]
15
[Run Now] button
16
[Analysis GUI] button
17
[Advanced] button
18
[Minimum Height]
19
[Z-Calibration] button
Chapter 3
The Screens of the N-STORM Software
Selects whether to display a screen of graph of the number of
bright points during acquisition of the dataset. To display the
graph, select this checkbox before clicking [Run Now]. Also,
when using Auto LP, select this checkbox.
Automatically adjusts the power of the laser so that the number
of bright points is constant in the imaging frames. The number
of bright points obtained when this button is clicked is regarded
as the reference value. (Can be used during image acquisition
only.)
Specifies the maximum of the adjustment range when the laser
power is automatically adjusted using the [Auto LP] button.
Starts image acquisition according to the settings configured in
the N-STORM control window.
Displays the N-STORM analysis window.
Displays detailed settings.
Specifies the threshold for the difference in intensities (between
the peak of each bright point and the background) that is used
when molecules are identified during real time analysis
(preview). If the intensity exceeds the threshold, the relevant
bright point is identified as being a molecule.
Performs calibration for 3D-STORM.
35
Description