Nikon Nivo Series Manual de interfaz de comunicación - Página 10
Navegue en línea o descargue pdf Manual de interfaz de comunicación para Instrumentos de medida Nikon Nivo Series. Nikon Nivo Series 34 páginas. Hardware
También para Nikon Nivo Series: Manual de instrucciones (32 páginas)
Nikon Communication Interface Manual
5-5. Format of commands
a) Remote control command
$MSR
Input
None
Output
None
Taking a distance measurement in Precise mode.
Remarks
Resolution and average of current settings are used in the measurement.
When the "Data output after MSR" feature is set to ON, the measured data is to be
output in the specified format.
$TRK
Input
None
Output
None
Taking a distance measurement in Normal mode.
Remarks
Resolution and average of current settings are used in the measurement.
When the "Data output after MSR" feature is set to ON, the measured data is to be
output in the specified format.
$HTR
Input
None
Output
None
Initiate continuous distance measurements. (AVE=0)
Remarks
Resolution is fixed to COARSE mode.
When the "Data output after MSR" feature is set to ON, the measured data is to be
output in the specified format.
$SHV
Input
None
Output
None
Remarks
Change display. (Same as [DSP] key on the instrument.)
$RHA
Input
None
Output
None
Remarks
Reset the horizontal angle to zero.
$BZR
Input
+
0 : Length (1 to 9)
Output
None
Remarks
$BZRn; Beep a tone for "Length" x 50 msec.
P10/34
Confidential