Option A - Spectrum View: The other indicator is that a broad peak shows up in the Spectrum plot
between 440-450nm. An example plot is shown below; keep in mind peak location and intensity may vary.
See our "Chemical Cheat Sheet" to learn about where and how each molecule fluoresces on the unit.
Arometrix recommends that you use the Spectrum plot for an absolute indicator of the Heads to main
body transition.
Option B - Wavelength View: This view will show that there is a big peak in the value of the wavelength
plot. An example plot is shown below. The Wavelength View is recommended to identify when the
distillation is transferring from Heads to Main Body.
Step 5) Perform Flask Transfer & Let Distillation System Equilibrate
1) Once you've entered Main Body, proceed as you typically would with a flask transfer.
a. It is advised that the Heads be allowed to distill for ~10 minutes before performing a flask
to ensure that no solvents contaminate the Main Body.
b. There is no need to turn off or hit the clear button on the FRACTION FINDER, unless you
want to.
2) After flask transfer, let the system come to its equilibrated point (let pressure and temperature
become relatively constant) before using the FRACTION FINDER
a. Keep in mind that the signal will be very low and all ambient lighting effects will greatly
increase
b. Equilibration time will depend on system size. A 5L SPD can take 5-10 mins. A 10L SPD can
be 10-15 minute. Mantle quality and pump quality are also factors for equilibration time.