Arthrex SynergyRF AR-9800 Manual de inicio rápido - Página 3
Navegue en línea o descargue pdf Manual de inicio rápido para Equipamiento médico Arthrex SynergyRF AR-9800. Arthrex SynergyRF AR-9800 3 páginas.
Button Activation
1. Ablation power setting
2. Ablate
3. Coag
Troubleshooting: Faults and Solutions
Fault
Unit does not power on
Accessory Apollo
probe does not work
RF®
Yellow fault message
Red fault message (Critical Failure)
To return or repair device: Contact Arthrex Product Surveillance at 1-866-267-9138.
Arthrex, Inc.
1370 Creekside Blvd.
Naples, FL 34108-1945 USA
Toll Free: 1 (800) 934-4404
www.arthrex.com
Arthrex Tech Support:
+1-888-420-9393
arthrex.com
© 2021 Arthrex, Inc. All rights reserved. LI1-00027-en-US_C
Solution
1. Check the main power plug and wall receptacle.
2. Check the main fuses per the User Guide instructions.
3. Call Arthrex Tech Support at 1-888-420-9393.
1. Disconnect and reconnect probe.
2. Turn off console, wait 30 seconds, repower console.
3. If applicable, disconnect and reconnect footswitch, replace if footswitch not
functioning.
4. Open new Apollo
probe.
RF
Refer to User Guide for proper disposition.
1. Make note of message displayed.
2. Call Arthrex Tech Support at 1-888-420-9393.
Arthrex GmbH
Erwin-Hielscher-Strasse 9
81249 München, Germany
Tel: +49 89 909005-0
www.arthrex.de
+49 89 90 90 05 8800
1
2
3