Arthur Grillo DA2000-K-S Manual de instalación y funcionamiento - Página 11
Navegue en línea o descargue pdf Manual de instalación y funcionamiento para Instrumentos de medida Arthur Grillo DA2000-K-S. Arthur Grillo DA2000-K-S 12 páginas. Differential pressure manometer
8.
Specification
Manometer
Measurment range:
Display:
Overload protection:
Static pressure:
Pressure connection:
Housing:
Ambient temperature:
Protection class:
Weight:
Tolerance:
Pressure switch
Adjustable switching ranges:
Tolerance on the upper and
lower switching point:
Switching capacity:
Durability:
Contact arrangement:
Analog output
Signal range:
Electrical connection:
Protection class:
0...100 Pa, 0...200 Pa, 0...500 Pa, 0...1000 Pa, 0...2000 Pa,
0...5000 Pa
Scale length 270° = approx. 250 mm
not less then 10-times measurment span
max. 0,2 bar
hose nozzle 8 mm ø
Ultramid/ABS, black, transparent cover
round outside 134 mm ø
-10...50 °C
IP 65
approx. 400 g
± 2 %
20...300 Pa, 30...400 Pa 50...500 Pa, 200...1000 Pa, 500...2500
Pa, 1000...5000 Pa
± 15 %
max. 1,5 A / 250 Vac
mechanical more then 10
normally closed contact - NC
normally open contact - NO
supply - COM
0...10 V, 3-wire-circuit
supply voltage 15...30 Vdc or 24 Vac
4...20 mA, 2-wire-circuit
supply voltage 12...30 Vdc
rectangle plug connector 16 mm or
color coded cable, approx. 50 cm long
IP54 - DA2000, DA2000-A, DA2000-K
IP66 - DA2000-S, DA2000-A-S, DA2000-K-S
Doc.-no: DA2000-01
TECHNICAL SPECIFICATIONS
cycles
6
issue: 02/2016
11