Maintenance log L2-Range and D2-Range
8. Cuspidor
Replace the biofilter in the cuspidor.
Glass filler: check drain cock at the retainer for leaks; if necessary, replace O-rings.
Remove any deposits of substances in water from the drain cock (mechanically or
in an ultrasonic bath).
Cuspidor bowl: check the drain of the cuspidor bowl for leaks; replace the O-seal if
necessary; clean the funnel.
Cuspidor bowl: check drain sieve (gold catcher) and replace.
With spitton valve 3: Perform a visual leak test.
Observe the manufacturer's (Dürr) maintenance instructions.
❯ Check compressed air supply ❯ Perform a functional test, clean the yellow coarse
filter or replace it if required.
With cuspidor valve 3: Perform a visual leak test.
Observe the manufacturer's (Dürr) maintenance instructions. ❯ Clean the float
monitor in the collection vessel. ❯ Clean the seal on the float sensor for ventilation
and replace if required, clean the yellow coarse filter or replace it if required.
Dürr rinsing unit II: Perform a visual leak test.
Check and clean the fine filter. Check the amount of dirt on the float monitor
(outlet) and to clean this where required.
With amalgam separator CAS1: Perform a visual leak test. Check function and
suction power: Observe the manufacturer's (Dürr) maintenance instructions.
❯ Cleaning of the suction unit in accordance with the operating instructions.
❯ Clean the float. ❯ Replace the bellows.
With amalgam separator CAS1: Perform a visual leak test. Observe the
manufacturer's (Dürr) maintenance instructions. ❯ Replace the rubber grommets on
the connections. ❯ Replace the float.
With amalgam separator CAS1: Perform a visual leak test. Observe the
manufacturer's (Dürr) maintenance instructions. ❯ Replace the centrifuge drum and
seal. ❯ Replace all O-rings (from the replacement parts kit) in the device. ❯ Replace
the rubber grommets on the connections. ❯ Replace the float.
L2-Range
Once a year
Bio filter maintenance set
D2-Range
Every 3
L2-Range
years
D2-Range
Every 3
L2-Range
years
D2-Range
L2-Range
Once a year
D2-Range
L2-Range
Once a year
D2-Range
Every 3
L2-Range
years
D2-Range
L2-Range
Once a year
D2-Range
L2-Range
Once a year
D2-Range
Every 3
L2-Range
Service kit (3-year interval)
years
D2-Range
Every 5
L2-Range
Service kit (5-year interval)
years
D2-Range
210620-W
(included in
maintenance
set 211390)
640003
(not included in
O-ring 8x1
maintenance set
211390)
640019
(not included in
O-ring seal 42x2,5
maintenance set
211390)
410081
(included in
Sieve
maintenance
set 211390)
0700-702-06E
Protective strainer
(order from
DÜRR Dental)
0700-702-06E
Protective strainer
(order from
DÜRR Dental)
-
-
7117-420-25E
Bellows
(order from
DÜRR Dental)
7117-980-32
(order from
DÜRR Dental)
7117-980-30
(order from
DÜRR Dental)
1
2
1
1
-
-
-
-
-
-
page 7