COMMUNICATION: Communication problem between
hand-held programming unit and the motor control unit
-> Hand-held programming unit, motor electronics or
spiral cable of the hand-held programming unit defective
CPM-DEVICE ERROR: General mistake in the motor
control unit
-> Replace motor electronics (see chapter 7.3) if
persistent
MOT. EN. TIMEOUT: Motor can not be activated
-> Check motor electronics and plug connection for
motor. Replace motor electronics (see chapter 7.3) and
spiral cable of the hand-held programming unit if
persistent
INVALID PARAMET.: Motor control unit received a
wrong parameter
-> Hand-held programming unit, motor electronics or
spiral cable of the hand-held programming unit defective
STOP RELEASE ERR: Stop line can not be released
-> Remove hand-held programming unit or spiral cable
(see chapter 7.2) if persistent
UNEXP. MOT. STOP: Motor stopped unexpectedly
-> Motor electronics or connecting line to motor defective
MOTOR DISABLED: Motor is disconnected
-> Motor electronics or connecting line to motor defective
MOTOR ERROR: Motor is not turning
-> Check motor cable and motor. Replace motor
electronics (see chapter 7.3) if persistent
RTC INIT ERROR: Hand-held programming unit defective
-> Replace hand-held programming unit (see chapter7.2)
if persistent
RTC COM ERROR: Communication problem with hand-
held programming unit
-> Replace hand-held programming unit (see chapter7.2)
if persistent
RTC ERROR: Hand-held programming unit defective
-> Replace hand-held programming unit (see chapter7.2)
if persistent
RANGE EXCEEDED: Potentiometer shows a value out
of range
-> Replace Potentiometer and/ or spiral cable
UNKNOWN CPM ERR.: Unknown error in the motor
control unit
-> Replace motor electronics (see chapter 7.3) if
persistent
UNDEFINED ERROR: Undefined error
-> Hand-held programming unit, motor electronics or
spiral cable of the hand-held programming unit defective
CPM 24V SUPPLY: 24V supply in motor control out of
range
-> Check power supply. Replace motor electronics (see
chapter 7.3) if persistent
CPM 3.3V SUPPLY: 3.3V supply in motor control out of
range
-> Replace motor electronics(see chapter 7.3) if persistent
HS 24V SUPPLY: 24V supply in hand-held programming
out of range
-> Replace hand-held programming unit or spiral cable if
persistent
HS 5V SUPPLY: 5V supply in hand-held programming unit
out of range
-> Replace hand-held programming unit if persistent
HS 3.3V SUPPLY: 3.3V supply in hand-held programming
unit out of range
-> Replace hand-held programming unit if persistent
3.2 Mechanics
The Threaded spindle is not interchangeable with
ARTROMOT®-K3 serial number < 3000.
Do not loosen the knurled handles completely for any
adjustment. For operation or transport, make sure that
the knurled handles are tight.
3.3 Others
Do not clean the casing or the support with grease
or oil.
Do not utilize any solvent for cleaning the
ARTROMOT®-K3.
3