Before mounting de-capper head:
4.
-Check needle for any damage: The needle should be sharp and in straight condition.
-Check spring stop: The spring should be in parallel line with the de-capper head.
To mount the de-capper head:
5.
Check that de-capper arm is put in outward position. See above "To remove de-
capper head".
Hold the head in a 90 degree left position and push de-capper head onto the
mounting arm. In pressed position, turn right (90 degree clockwise) and release.
The head should now be in locked normal position.
Press Red button once, then Start to let Pluggo go to automatic stop.
To mount chute: Hold chute with spring downwards and press onto holder.
6.
Close cover and check no interfering with chute.
7.
Pluggo is now ready for operation.