OPERATION FUNCTIONS:
1. THREE-WAY AIR/WATER SYRINGE -
Depress the right button for air operation,
and the left button for water operation.
Depressing both buttons will create a mist.
The syringe features quick-change
autoclavable tips: To remove a tip, press
on the locking collar surrounding the tip
socket and pull the used tip straight out of
the socket (see Fig.6). To insert a new tip,
press locking collar and push tip into
socket as far as it will go. Release collar
and gently tug on tip before using to
ensure that tip is securely locked into
socket.
Syringe Tip Sterilization:
1) Remove contaminated syringe tip.
2) Remove all visible signs of contamination
before autoclaving.
3) Autoclave tip at 132° C (270° F) for ten
minutes.
4) Sterilize between each patient use.
Figure 6
NOTE: Since only the tips can be
autoclaved, it is recommended that the
air/water syringe be bagged with a
disposable, single-use plastic sleeve
between each patient use.
2. HANDPIECE CONTROLS (Fig. 7) - The
Aseptico ADU-20B includes two
handpieces mounted to the front bar.
These are fully automatic. By placing a
handpiece into its proper holder, the
4
DEPRESS
LOCKING COLLAR
pressure is automatically shut off. When a
handpiece is removed from it's holder,
pressure is automatically opened and
ready for use with foot control.
The amount of air pressure supplied to
each handpiece may be adjusted by
turning the flow control for that handpiece
on the control block (See page 8).
Figure 7
3. WET/DRY FOOT CONTROL (Fig. 8) - The
foot control activates the handpieces and
water coolant spray.
Position the foot control on the floor. For
handpiece operation, apply foot pressure
to any part of the center disk. The
handpiece must be removed from its
holder before operation can begin.
WATER
APPLY
PRESSURE TO
ACTIVATE
HANDPIECE
Figure 8 - Foot Control
3-WAY
AIR/WATER
SYRINGE
LOWSPEED
CONTROL
HIGHSPEED CONTROL
WATER
ON
OFF