Your new Aseptico ADU-20B CompriCart II delivery system is
one of the finest portable carts available to the dental profession.
The system is designed for portable operatory use and features
a three-way air/water syringe, automatic highspeed and
lowspeed handpiece controls, high volume evacuator and saliva
ejector. A pressurized water tank supplies water for handpiece
coolant and syringe. The system is powered by a single
connection to compressed air via the air supply line.
Congratulations!
This Aseptico ADU-20B system is engineered to provide many
years of reliable service. Please read the instructions provided
in this manual to receive the best and longest service from your
Aseptico equipment.
Separate manuals may be provided to cover the operation and
maintenance of hand-held pieces or other accessories for your
unit.
PACKAGE CONTENTS
•
Delivery Unit Cart
•
TA-90D Three-Way Syringe
•
TA-1 Syringe Tip Autoclavable 3-way
•
AA-25 Vacuum Saliva Ejector
•
AA-21 Vacuum High Volume AVS
•
2-Liter Water Reservoir
•
AA-42W Foot Control
•
AA-50 Instrument Tray
•
NWS-7i Water Sys Self-cont Installed
•
AHP-101 Handpiece 1:1 Straight Low Speed Doriot
(Optional)
•
AA-19LED-04 Fiber Optic 1 Hp Illumination System Led
Straight Gray (Tubing 6 Ft) (Optional)
•
AA-19EGEN-04T6 6-Pin Self-Generating Fiber Optic
Handpiece Connection (Optional)
•
AA-290CK Central Vacuum Kit (Optional)
•
AWS-1T Aseptiwater Tap Water option (Optional)
•
AWS-1C Aseptiwater Cavitron® Connect (Optional)
Cavitron® is a registered trademark of DENTSPLY International Inc.
SETTING UP THE SYSTEM:
1.
Place the ADU-20B CompriCart II in an upright position on a
flat, stable surface. Loosen the center column retaining knob
and extend the telescoping upper portion. Tighten the knob
at the desired height (Fig. 1).
2.
Check to make sure the syringe, highspeed and lowspeed
handpieces are hanging in their proper holder. Locate the
foot control on the floor (Fig. 1).
Removable
Instrument
Tray
Three-Way
Syringe
Highspeed
Handpiece
Control
Lowspeed
Handpiece
Control
Telescoping
Center Column
with Retaining
Knob
Foot Control
Figure 1 - ADU-20B System
3.
The ADU-20B unit must be connected to a dry, filtered,
70-100 psi air source. Simply extend the 1/4" grey air
supply line from the ADU-20B cart and connect the quick-
disconnect into the compressed air source (Fig. 2).
Figure 2 - Air Supply Connector
4.
Remove the high volume vacuum system, saliva ejector
system, and High-Density Polyethylene (HDPE) water
supply tank from packing and install on proper quick
disconnects on the delivery head (Fig. 1).
5.
To install the vacuum bottle, press on locking ring tab, push
bottle into quick coupler, and release tab (Fig. 3a).
A.
Hang the vacuum head in the holder at the right hand
Figure 3b end of the front holder rack (Fig. 3b).
Figure 3a
Quick Disconnect
Locking Ring
Vacuum Bottle
Tab
6.
To install the saliva ejector system, locate the quick
disconnect at the rear of the delivery head (Fig. 4a). Note
manual release lever on disconnect. Insert metal fitting on
Pressure Gage
AA-21 High
Volume
Evacuator
Assembly
2-Liter HDPE
Water Supply
Saliva Ejector
Connect to
70-100 PSI Air
Source
Figure 3b
Handpiece
Tank
System
3.