TROUBLESHOOTING:

Problem:
"On" Indicator does not light
when turned on:
"On" Indicator lights when turned
on, but handpiece does not turn:
"On" Indicator dims and goes
out, and/or motor slows and
stops working.
Correction:
• Check console to power connection.
• Check that voltage is proper voltage.
• Check fuse. If blown, replace with 5x20mm
0.5A/250V fast-acting fuse.
• Depress foot pedal (if used).
• Check foot pedal connection (if used).
• Check motor plug connection.
• Test for possible bad handpiece by removing
handpiece from motor and then running motor by
itself.
• Check that Hand/Foot Selector switch is placed
fully in either the "Hand" or "Foot" positions
and isn't set halfway between.
• Check that a bit is properly seated in the
handpiece and the collet is closed.
• Test for possible motor/cable short by removing
motor cable from motor receptacle.
• Test for possible bad bearings in the motor by
removing the handpiece from the motor then
running motor by itself.