Following are the functions of the buttons when in normal operating mode.
In normal operation, pressing the RESET button resets the torque readings to zero in
either the PEAK or FIRST PEAK mode.
Note! In TRACK MODE pressing the RESET button will take you into the MAIN MENU.
Press the button again to ESCAPE back to the previous screen and press the MODE
button to return to PEAK or FIRST PEAK Mode.
Button #1: MENU, RESET/ESCAPE
In normal operation this is the reset button. Press to reset the display to zero
prior to taking a torque reading.
Button #2: MODE
In normal operation, the MODE button is used to change the MODE setting of the DTT
meter. Each time the button is pressed the mode will change from TRACK mode to
PEAK, or from PEAK to FIRST PEAK, or from FIRST PEAK to TRACK.
Button #3: UNIT
In normal operation, this button changes the torque units of the meter. Pressing the
button will scroll through the 9 choices. Kgf-cm, gf-cm, Kgf-m, mN-m, cN-m, N-m,
ozf-in, lbf-in, lbf-ft, in this order. Do not attempt to back up by pressing the DOWN
button; you will only change the mode.
Button #4: MEM (Memory)
In normal operation, pressing the MEM button enters the torque reading into the DTT
memory.
5)

OVERLOAD CAPACITY

Do not deliberately exceed the capacity of the unit. The maximum capacity of
the DTT meter can be found on the spacer ring surrounding the transducer
The overload capacity of a DTT meter is 200% of its maximum load capacity. If the
meter is taken over capacity; a warning tone will sound, the word "OVERLOAD" will
flash on the LCD and the unit will LOCK until the RESET button is pressed.
Immediately stop applying torque when the overload warnings are observed.
A unit subjected to overload, or subjected to harsh conditions such as use with an
impact tool will not be covered under warranty.
6)

PROCEDURE FOR MANUAL TORQUE DRIVERS

Attach a proper adapter to connect the driver to the 3/8" female square. NOTE IT IS
NOT RECCOMMENDED TO USE THE POWER TOOL ADAPTORS TO CALIBRATE
HAND TORQUE TOOLS
Select FIRST PEAK mode using the "MODE" button. Press the "RESET / ESC"
button if the display is not at 0.
Turn the driver or wrench clockwise or counterclockwise until the "break" point is
reached. In FIRST PEAK no other reading can then be displayed until the reset button
is pressed.
CAUTION!
5