WARNING
When using hazardous chemicals, always use
protective glasses and other suitable personal
protection equipment.
Installation
The instrument is delivered pre-installed in ÄKTAprocess system
with all electrical connections prepared inside the electronics
cabinet. The conductivity cells and the pH electrode flow cell holder
are also mounted in proper positions.
WARNING
The electronics cabinet of the ÄKTAprocess
system must only be opened by personnel
authorized by Cytiva.
Operation
When the system is powered, the instrument is simultaneously
switched on. The monitor can be used immediately, but the full
specifications are not reached until after a warm-up period of
15 minutes.
The pH/C-902 Monitor has no user interface for the operator. It is
controlled entirely via UNICORN control software (version 5.0 or
later). The pH and conductivity values, as well as the temperature, are
displayed in the System Control module of the UNICORN control
software.
Note: The measured temperature is the temperature in the
conductivity flow cell, which can differ from the ambient
temperature.
Preparing the instrument
The following settings and calibrations may be necessary to perform
as described below:
•
Set up conductivity temperature compensation
•
Calibrate the conductivity cells
•
Prepare and insert the pH electrode
•
Calibrate the pH electrode
•
Set up pH temperature compensation
Setting up the conductivity temperature compensation
The conductivity in a buffer is temperature dependent. Temperature
compensation should be used when relating conductivity to
concentration and/or comparing conductivity values. The
compensation consists of a compensation factor together with a
reference temperature. All conductivity values will then
automatically be converted to the set reference temperature.
2
In the UNICORN System Control module:
Step
Action
Select System:Settings.
1
Result: The Instructions dialog box opens.
Select Monitors and click Cond_CIR101 or Cond_CIR102.
2
3
Enter a temperature compensation value between 0 and
9.99% in the CompFactor box.
Click OK.
4
The factor is expressed in percentage increase of conductivity per ℃
increase in temperature. If the temperature compensation factor is
unknown, a general approximate value of 2% can be set for many
common salt buffers.
Check the OFF-box in the Instructions dialog box for no
temperature compensation.
Calibrating the conductivity cell
Normally it is not necessary to adjust the cell constant as the flow cell
is pre-calibrated on delivery. The conductivity flow cell constant is
shown on the packaging and can be entered when the cell is
changed. Adjustment is only necessary when replacing the
conductivity cell with a flow cell whose constant is unknown.
However, it is also recommended that the flow cell is recalibrated
after cleaning.
The cell constant is adjusted in the UNICORN System Control
module.
Setting the constant for a new conductivity cell
The following instruction is applicable for a conductivity cell where
the constant value is known:
Step
Action
1
Read the constant value from the packaging.
2
Selec System:Calibrate.
Result: The Calibration dialog box opens.
3
In the Monitor box, select CIT101_Cel or CIT102_Cel.
4
Type the constant value in the Reference Value 1 box.
Click the Read Value 1 button.
5
Result: The measured value is displayed in the Measured
value box.
6
Click the Close button to exit the dialog box.
Calibrating the conductivity cell
Note: The conductivity temperature compensation must not be
activated when the cell constant is adjusted. Set the
compensation to OFF in the Instructions dialog box.
Step
Action
1
Prepare a calibration solution of 1.00 M NaCl, 58.44 g/l. Let the
solution rest until it is at room temperature. This is important
for exact measurements.
2
Fill the flow cell completely with the calibration solution by
pumping at least 15 ml through the cell with a syringe.
pH/C-902 Instruction 14003958 AB